Request a quote now, contact your distributor or our sales offices or click here.
Product
MSC
Multi-signal and multi-function portable smart calibrator
MSC (Multifunction Smart Calibrator) is a flexible and universal instrument for the maintenance and testing of process sensors and meters. With an accuracy class of better than 0.05% for each type of input/output, MSC offers measurement and signal generation: analog, digital, temperature sensor and load cell. This powerful calibrator offers advanced features such as automatic ramp generation (single or loop) and data logging with instant data sharing/exporting or an entire log session in .CSV format.
Equipped with a rechargeable lithium-polymer battery, MSC is able to power external devices or sensors and can be used without continuous power supply with the autonomy of up to 20 hours. With a reading capacity of up to 100,000 measurements and the full support of the LabVIEW libraries, it is suitable for professional and industrial use by PLC programmers, maintenance technicians, technical assistance companies, measurement laboratories, device control and calibration activities, electronics laboratories, electromechanical workshops, and quality control activities.
The device is supplied with a calibration report useful to verify its metrological characteristics and to validate the correctness and accuracy of the measured/generated signal during calibration sessions.
Parameter configuration and data management can be done from a Windows desktop application (via micro USB cable) or mobile app (for iOS and Android devices via Bluetooth connection).
Request a quote now, contact our sales offices or click here.
Equipped with a rechargeable lithium-polymer battery, MSC is able to power external devices or sensors and can be used without continuous power supply with the autonomy of up to 20 hours. With a reading capacity of up to 100,000 measurements and the full support of the LabVIEW libraries, it is suitable for professional and industrial use by PLC programmers, maintenance technicians, technical assistance companies, measurement laboratories, device control and calibration activities, electronics laboratories, electromechanical workshops, and quality control activities.
The device is supplied with a calibration report useful to verify its metrological characteristics and to validate the correctness and accuracy of the measured/generated signal during calibration sessions.
Parameter configuration and data management can be done from a Windows desktop application (via micro USB cable) or mobile app (for iOS and Android devices via Bluetooth connection).
Request a quote now, contact our sales offices or click here.
Contact us for
Support & Info
TECHNICAL DATA
Battery power supply | Lithium Polymer (LiPo) battery 3.400 mAh |
Lifetime | Lifetime 8 hours (min@max load), 20 hours (max) |
AC Power supply | 230 Vac with standard USB battery charger |
Status LEDs | power supply communication error pairing BT datalogger on (TBD) battery status |
Insulation | Battery powered instrument, intrinsically isolated |
Overvoltage protection | 230 Vac max without permanent damages (TBC) |
Rejection | 50/60 Hz |
Sampling rate | 10 Hz |
Operating modes | Signal measurement Signal generation (with ramps) Datalogging |
Buzzer | for overload reporting and inability to simulate the requested value |
Standard bushings | Nr. 4 test leads 4mm |
Thermocouple connection | Mini plug (7,9 mm) for TC signal measurement or generation |
MicroUSB functions | power supply FW upgrade ModBUS communication (virtual COM) |
Wireless communication | Bluetooth Low Energy 4.1 for iOS or Android device connection |
Mobile app | MSC by SENECA. Language available: ITA, ENG, DEU, FRA, ESP |
Operating temperature | -20..50°C (without charging), 0-45°C during charge |
Error display | Overvoltage (V > 28 V) Under voltage (V < -0,2 V) Over current (I > 24 mA) Under current (I < -0,1 mA) Resistance < 18 Ohm Resistance > 400 Ohm Load cell fail supply Blink value (voltage/current/resistance/bridge generation failure) |
Humidity | 30..90 % non condensing |
Dimension (WxHxD) | 88 x 147 x 25 mm |
Weight | 330 g |
Protection degree | IP20 |
Equipment | Connection cables (4), microUSB/USB cable, battery charger |
Calibration certificate | On demand |
Standard | CE |
Norms | EN61326-1; EN61010-1 |
SIGNAL MEASUREMENT | |
Current | 0..24 mA active & passive; protection ± 28 V |
Voltage (V) | 0,0÷27 V |
Voltage | -10mV÷+90mV |
Thermocouple | Type J,K ,T, E, N, R, S, B, L |
Thermoresistors (2,3,4 wires) | Pt100, Pt500, Pt1000, Cu50, Cu100, Ni100, Ni120 |
Load cell | 350 Ohm; -0,2÷+2,4mV/V |
Pulse | max count 1.000 Hz |
Frequency signal | 0,1..1.000 Hz |
Accuracy | 0,03%; 0,04% for current |
Resolution | 1 µA; 1 mV; 5 µV; 0,1°C; 0,1uV/V |
SIGNAL GENERATION | |
Current | 0,1..24 mA active & passive; protection ± 28 V |
Voltage (V) | 0,1÷24 V |
Voltage (mV) | -10mV÷+90mV |
Thermocouple | Tipo J,K ,T, E, N, R,S, B, L |
Thermoresistors (2 wires) | Pt100, Pt500, Pt1000, Cu50, Cu100, Ni100, Ni120 |
Load cell | 350 Ohm; -0,2..+2,4mV/V |
Pulse | min 0,5 ms (1..24V) pulses number settable |
Frequency signal | 0,1..1.000 Hz |
Accuracy | 0,03%; 0,04% for current |
Resolution | 1 µA; 1 mV; 5µV; 0,1°C; 0,02 Ohm; 0,1 uV/V; |
RAMP FUNCTIONS | |
Signals | Current/Voltage/TC/RTD/Load cell |
Functions | single or in Loop |
Type | MAX 9 segments, ramp resolution 100 ms, minimum ramp 1 second |
DATALOGGING | |
Datalogger | Yes |
Sampling rate | >500 ms |