Release notes Firmware SW01050-1.0.0.42 Supported product: MSC Fix: - the battery was charged only once then stopped charging until the instrument restarts, now it is charged correctly as soon as it starts to discharge - on TC measurement, the secondary measurement did not report a consistent usable data, now measures the TC voltage value without cold junction correction - datalogger did not restart with zero numbering when memory was deleted, the downloaded csv file now restarts with the correct numbering of the individual readings