The USB-ISO accessory is mandatory for certain use cases:
- in case you need to connect several devices (e.g. several MSC calibrators) to the same workstation (a USB-ISO isolator is required for each device)
- in case your MSC calibrator needs to be connected to a grounded circuit
- in case several MSC calibrators are to be connected to the same circuit at the same time
TECHNICAL DATA
| Battery power supply | Lithium Polymer (LiPo) battery 3.400 mAh |
| Lifetime | Lifetime 8 hours (min@max load), 20 hours (max) |
| AC Power supply | 230 Vac with standard USB battery charger |
| Status LEDs | power supply communication error pairing BT datalogger on (TBD) battery status |
| Insulation | Battery powered instrument, intrinsically isolated |
| Overvoltage protection | 230 Vac max without permanent damages (TBC) |
| Rejection | 50/60 Hz |
| Sampling rate | 10 Hz |
| Operating modes | Signal measurement Signal generation (with ramps) Datalogging |
| Buzzer | for overload reporting and inability to simulate the requested value |
| Standard bushings | Nr. 4 test leads 4mm |
| Thermocouple connection | Mini plug (7,9 mm) for TC signal measurement or generation |
| MicroUSB functions | power supply FW upgrade ModBUS communication (virtual COM) |
| Wireless communication | Bluetooth Low Energy 4.1 for iOS or Android device connection |
| Mobile app | MSC by SENECA. Language available: ITA, ENG, DEU, FRA, ESP |
| Operating temperature | -20..50°C (without charging), 0-45°C during charge |
| Error display | Overvoltage (V > 28 V) Under voltage (V < -0,2 V) Over current (I > 24 mA) Under current (I < -0,1 mA) Resistance < 18 Ohm Resistance > 400 Ohm Load cell fail supply Blink value (voltage/current/resistance/bridge generation failure) |
| Humidity | 30..90 % non condensing |
| Dimension (WxHxD) | 88 x 147 x 25 mm |
| Weight | 330 g |
| Protection degree | IP20 |
| Equipment | Connection cables (4), microUSB/USB cable, battery charger |
| Calibration certificate | On demand |
| Standard | CE |
| Norms | EN61326-1; EN61010-1 |
| SIGNAL MEASUREMENT | |
| Current | 0..24 mA active & passive; protection ± 28 V |
| Voltage (V) | 0,0÷27 V |
| Voltage | -10mV÷+90mV |
| Thermocouple | Type J,K ,T, E, N, R, S, B, L |
| Thermoresistors (2,3,4 wires) | Pt100, Pt500, Pt1000, Cu50, Cu100, Ni100, Ni120 |
| Load cell | 350 Ohm; -0,2÷+2,4mV/V |
| Pulse | max count 1.000 Hz |
| Frequency signal | 0,1..1.000 Hz |
| Accuracy | 0,03%; 0,04% for current |
| Resolution | 1 µA; 1 mV; 5 µV; 0,1°C; 0,1uV/V |
| SIGNAL GENERATION | |
| Current | 0,1..24 mA active & passive; protection ± 28 V |
| Voltage (V) | 0,1÷24 V |
| Voltage (mV) | -10mV÷+90mV |
| Thermocouple | Tipo J,K ,T, E, N, R,S, B, L |
| Thermoresistors (2 wires) | Pt100, Pt500, Pt1000, Cu50, Cu100, Ni100, Ni120 |
| Load cell | 350 Ohm; -0,2..+2,4mV/V |
| Pulse | min 0,5 ms (1..24V) pulses number settable |
| Frequency signal | 0,1..1.000 Hz |
| Accuracy | 0,03%; 0,04% for current |
| Resolution | 1 µA; 1 mV; 5µV; 0,1°C; 0,02 Ohm; 0,1 uV/V; |
| RAMP FUNCTIONS | |
| Signals | Current/Voltage/TC/RTD/Load cell |
| Functions | single or in Loop |
| Type | MAX 9 segments, ramp resolution 100 ms, minimum ramp 1 second |
| DATALOGGING | |
| Datalogger | Yes |
| Sampling rate | >500 ms |