The USB-ISO accessory is mandatory in certain applications:
- when several devices (e.g. multiple MSC calibrators) need to be connected to the same workstation (one USB-ISO isolator required for each device)
- when your MSC calibrator needs to be connected to a grounded circuit
- when several MSC calibrators need to be connected simultaneously to the same circuit
TECHNICAL DATA
| Battery power supply | 3400 mAh Lithium Polymer (LiPo) battery |
| Autonomy | 8-hour operating time (minimum @ max load), 20 hours (max) |
| Mains power supply | 230 Vac mains supply with standard power adapter and 5 V 0,5 A USB socket |
| Status LEDs | power-on communication error BT pairing datalogger on (future) battery status |
| Isolation | Battery-powered, intrinsically isolated instrument. No isolation from USB port |
| Overvoltage protection | 230 Vac max without permanent damage (TBC) |
| Rejection | 50/60 Hz |
| Sampling freq. | 10 Hz |
| Operating Modes | Meter Generator (with ramp functions) Datalogger |
| Buzzer | Buzzer for signalling overload and inability to simulate the requested value. |
| Standard sockets | No. 4 4mm sockets |
| Thermocouple connection | Mini plug (7,9 mm) for thermocouple measurement and simulation |
| MicroUSB port functions | power supply FW update ModBUS communication (virtual com) |
| Wireless communication | Bluetooth Low Energy 4.1 to Andriod or Ios Smartphones and Tablets |
| Languages available for mobile APP | Italian, English, German, French, Spanish |
| Operating temperature | -20..50°C (not charging), 0-45°C while charging |
| Error indications | Overvoltage Voltage reading above 28 V Under voltage Voltage reading below -0,2 V Over current Current reading above 24 mA Under current Current reading below -0,1 mA Resistance below 18 Ohm Resistance above 400 Ohm Load cell not powered correctly Flashing value Voltage / current / resistance / bridge generation failed |
| Humidity | 30..90% non-condensing |
| Dimensions (WxHxD) | 88 x 147 x 25 mm |
| Weight | 330 g |
| Protection rating | IP20 |
| Included | Connection leads (4), USB-microUSB cable, mains battery charger |
| Calibration certificate | On request |
| Approval | CE |
| Standards | EN61326-1; EN61010-1 |
| MEASUREMENT FUNCTIONS | |
| Current | 0..24 mA active and passive; protection ± 28 V |
| Voltage (V) | 0,0÷27 V |
| Voltage | -10mV÷+90mV |
| Thermocouple | Type J,K ,T, E, N, R, S, B, L |
| RTDs (2,3,4 wires) | Pt100, Pt500, Pt1000, Cu50, Cu100, Ni100, Ni120 |
| Load cell | 350 Ohm; -0,2÷+2,4mV/V |
| Pulse | Maximum count 1000 Hz |
| Frequency | 0,1..1000 Hz |
| Measurement accuracy | 0,03% basic, 0,04% for current |
| Measurement resolution | 1 µA; 1 mV; 5 µV; 0,1°C; 0,1uV/V |
| GENERATION FUNCTIONS | |
| Current | 0,1..24 mA active and passive; protection ± 28 V |
| Voltage (V) | 0,1÷24 V |
| Voltage (mV) | -10mV÷+90mV |
| Thermocouple | Type J,K ,T, E, N, R,S, B, L |
| RTDs (2 wires) | Pt100, Pt500, Pt1000, Cu50, Cu100, Ni100, Ni120 |
| Load cell | 350 Ohm; -0,2..+2,4mV/V |
| Pulse | min 0,5 ms (1..24V), configurable number of pulses |
| Frequency | 0,1..1000 Hz |
| Generation accuracy | 0,03% basic, 0,04% for current |
| Generation resolution | 1 µA; 1 mV; 5µV; 0,1°C; 0,02 Ohm; 0,1 uV/V; |
| RAMP FUNCTION | |
| Signal | Current/Voltage/TC/RTD/Load cell |
| Functions | Single or Loop |
| Type | Maximum 9 segments, 100ms ramp resolution, minimum ramp 1 second |
| DATALOGGER | |
| datalogger | Yes |
| sampling time | >500 ms |
Manuels, prospectus
Tools
LabVIEW driver
SHA256 Checksum:
89c9e06d4d4a694b8712dc36437104e8f2480796d68734b38381c98d570e6ba8
SHA256 Checksum:
89c9e06d4d4a694b8712dc36437104e8f2480796d68734b38381c98d570e6ba8
Firmwares
Conformité des produits
Produits alternatifs
Versions du produit
Accessoires
CU-A-MICROB
USB cable plug USB-A - MicroUSB-B - 5P ( KIT - USB , MY2 , Z109REGBP )
Disponibilité à confirmer, nous contacter ?
FR

Ecrivez-nous pour