The USB-ISO accessory is mandatory for certain use cases:
- in case you need to connect several devices (e.g. several MSC calibrators) to the same workstation (a USB-ISO isolator is required for each device)
- in case your MSC calibrator needs to be connected to a grounded circuit
- in case several MSC calibrators are to be connected to the same circuit at the same time
TECHNICAL DATA
Battery power supply | Lithium Polymer (LiPo) battery 3.400 mAh |
Lifetime | Lifetime 8 hours (min@max load), 20 hours (max) |
AC Power supply | 230 Vac with standard USB battery charger |
Status LEDs | power supply communication error pairing BT datalogger on (TBD) battery status |
Insulation | Battery powered instrument, intrinsically isolated |
Overvoltage protection | 230 Vac max without permanent damages (TBC) |
Rejection | 50/60 Hz |
Sampling rate | 10 Hz |
Operating modes | Signal measurement Signal generation (with ramps) Datalogging |
Buzzer | for overload reporting and inability to simulate the requested value |
Standard bushings | Nr. 4 test leads 4mm |
Thermocouple connection | Mini plug (7,9 mm) for TC signal measurement or generation |
MicroUSB functions | power supply FW upgrade ModBUS communication (virtual COM) |
Wireless communication | Bluetooth Low Energy 4.1 for iOS or Android device connection |
Mobile app | MSC by SENECA. Language available: ITA, ENG, DEU, FRA, ESP |
Operating temperature | -20..50°C (without charging), 0-45°C during charge |
Error display | Overvoltage (V > 28 V) Under voltage (V < -0,2 V) Over current (I > 24 mA) Under current (I < -0,1 mA) Resistance < 18 Ohm Resistance > 400 Ohm Load cell fail supply Blink value (voltage/current/resistance/bridge generation failure) |
Humidity | 30..90 % non condensing |
Dimension (WxHxD) | 88 x 147 x 25 mm |
Weight | 330 g |
Protection degree | IP20 |
Equipment | Connection cables (4), microUSB/USB cable, battery charger |
Calibration certificate | On demand |
Standard | CE |
Norms | EN61326-1; EN61010-1 |
SIGNAL MEASUREMENT | |
Current | 0..24 mA active & passive; protection ± 28 V |
Voltage (V) | 0,0÷27 V |
Voltage | -10mV÷+90mV |
Thermocouple | Type J,K ,T, E, N, R, S, B, L |
Thermoresistors (2,3,4 wires) | Pt100, Pt500, Pt1000, Cu50, Cu100, Ni100, Ni120 |
Load cell | 350 Ohm; -0,2÷+2,4mV/V |
Pulse | max count 1.000 Hz |
Frequency signal | 0,1..1.000 Hz |
Accuracy | 0,03%; 0,04% for current |
Resolution | 1 µA; 1 mV; 5 µV; 0,1°C; 0,1uV/V |
SIGNAL GENERATION | |
Current | 0,1..24 mA active & passive; protection ± 28 V |
Voltage (V) | 0,1÷24 V |
Voltage (mV) | -10mV÷+90mV |
Thermocouple | Tipo J,K ,T, E, N, R,S, B, L |
Thermoresistors (2 wires) | Pt100, Pt500, Pt1000, Cu50, Cu100, Ni100, Ni120 |
Load cell | 350 Ohm; -0,2..+2,4mV/V |
Pulse | min 0,5 ms (1..24V) pulses number settable |
Frequency signal | 0,1..1.000 Hz |
Accuracy | 0,03%; 0,04% for current |
Resolution | 1 µA; 1 mV; 5µV; 0,1°C; 0,02 Ohm; 0,1 uV/V; |
RAMP FUNCTIONS | |
Signals | Current/Voltage/TC/RTD/Load cell |
Functions | single or in Loop |
Type | MAX 9 segments, ramp resolution 100 ms, minimum ramp 1 second |
DATALOGGING | |
Datalogger | Yes |
Sampling rate | >500 ms |
Manuels, prospectus
Tools
LabVIEW driver
SHA256 Checksum:
89c9e06d4d4a694b8712dc36437104e8f2480796d68734b38381c98d570e6ba8
SHA256 Checksum:
89c9e06d4d4a694b8712dc36437104e8f2480796d68734b38381c98d570e6ba8

Firmwares
Certifications
Produits alternatifs
Dear Customer, please note that our offices will be closed from August 8 to August 24 due to the summer holidays. The last shipping day before the break is Wednesday, August 6. We kindly ask you to place any orders by August 5. Orders received after this deadline will be processed starting from Monday, August 25, when regular operations resume. Thank you for your cooperation.
Versions du produit
Accessoires
CU-A-MICROB
USB cable plug USB-A - MicroUSB-B - 5P ( KIT - USB , MY2 , Z109REGBP )
Disponible en quantité limitée ?